After setting the "Compare" condition, you can enter the identifier value by setting the
state high, low, or X (do not care) for each single bit. Alternatively, you can enter a hex-
adecimal value for each half byte.
The settings are the same as for the setup of the identifier trigger, see also "Identifier
condition" on page 304.
If a label list with node names was loaded and applied in the bus configuration, you
can select the node name from the list instead of entering the numeric identifier. The
instrument triggers on the identifier of the selected node.
Remote command:
SEARch:PROTocol:LIN:ICONdition on page 665
SEARch:PROTocol:LIN:IDENtifier on page 666
Data condition
Settings to define the data pattern to be searched, if "Event" = "Id & Error" is selected.
After setting the "Data" length and the "Compare" condition, you can enter the data
value by setting the state high, low, or X (do not care) for each single bit. Alternatively,
you can enter a hexadecimal value for each half byte.
The settings are the same as for the setup of the data trigger, see also "Data condition"
on page 305.
Remote command:
SEARch:PROTocol:LIN:DLENgth on page 666
SEARch:PROTocol:LIN:DCONdition on page 666
SEARch:PROTocol:LIN:DATA on page 666
13.6.6 LIN label list
For general information on label lists, see Chapter 13.1.5, "Label list", on page 257.
Label lists are protocol-specific. Label lists for LIN are available in CSV and PTT for-
mat.
A LIN label file contains two values for each identifier:
●
Identifier value
●
Symbolic name for the identifier
Example of a LIN PTT file
# ----------------------------------------------------------------------------
@FILE_VERSION = 1.0
@PROTOCOL_NAME = lin
# ----------------------------------------------------------------------------
# Labels for LIN protocol
# Column order: Identifier, Label
# ----------------------------------------------------------------------------
# Labels for standard addresses
0x3F,Temperature
1Ch,Left brake
20h,Right brake
LIN (option R&S