Waveform setup
R&S
®
RTM3000
62User Manual 1335.9090.02 ─ 09
5.4.2 Probe settings for probes with BNC connector
For passive probes, which are connected with a BNC connector, you set the probe
attenuation and the unit, and you can start an adjustment procedure for the probe. All
settings are channel-specific. When you set the attenuation, you can select a prede-
fined factor, for example "10:1", or enter a user-defined value.
If AC coupling is set, the attenuation of passive probes has no effect, and voltage is
applied to the instrument with factor 1:1. Observe the voltage limits, otherwise you can
damage the instrument.
Access: [Ch <n>] > "Probe" (scroll down). Or: short menu > "Probe"
User
If default values do not fit, you can enter an arbitrary attenuation factor. The vertical
scaling and measured values are multiplied by this factor so that the displayed values
are equal to the undivided measured signal values.
Remote command:
PROBe<m>:SETup:ATTenuation:MANual on page 442
PROBe<m>:SETup:GAIN:MANual on page 443
Unit
Selects the unit that the probe can measure.
V - for voltage measurements
A - for current measurements
Remote command:
PROBe<m>:SETup:ATTenuation:UNIT on page 442
PROBe<m>:SETup:GAIN:UNIT on page 442
Probe Adjust
Starts the probe adjustment procedure. A wizard explains the adjustment step by step.
Probes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals