Remote control commands
R&S
®
RTM3000
711User Manual 1335.9090.02 ─ 09
WORD
Search for a word type.
Set the word type with SEARch:PROTocol:ARINc:WORD[:
TYPE].
ERROr
Search for errors of one or more error types.
Set the error types with SEARch:PROTocol:ARINc:ERRor.
LABel
Searches for label.
Set the label with SEARch:PROTocol:ARINc:LABel:
CONDition, SEARch:PROTocol:ARINc:LABel:MAXimum
and SEARch:PROTocol:ARINc:LABel:MINimum
LDATa
Searches for label and data.
Set the label with SEARch:PROTocol:ARINc:LABel:
CONDition, SEARch:PROTocol:ARINc:LABel:MAXimum
and SEARch:PROTocol:ARINc:LABel:MINimum
Set the data with SEARch:PROTocol:ARINc:DATA:
CONDition, SEARch:PROTocol:ARINc:DATA:MAXimum,
SEARch:PROTocol:ARINc:DATA:MINimum, SEARch:
PROTocol:ARINc:DATA:OFFSet, SEARch:PROTocol:
ARINc:DATA:SIZE, SEARch:PROTocol:ARINc:SDI and
SEARch:PROTocol:ARINc:SSM.
SEARch:PROTocol:ARINc:DATA:CONDition
Sets the comparison condition for data: If the pattern contains at least one X (don't
care), you can trigger on values equal or not equal to the specified value. If the pattern
contains only 0 and 1, you can also trigger on a range greater than or lower than the
specified value.
The command is relevant if SEARch:PROTocol:ARINc:CONDition is set to LDATa.
Parameters:
<DataCondition> EQUal | NEQual | GTHan | GEQual | LEQual | LTHan | WITHin |
OUTSide
SEARch:PROTocol:ARINc:DATA:MAXimum
Searches for a maximum value of the data if SEARch:PROTocol:ARINc:DATA:
CONDition is set to INRange or OORange.
The command is relevant if SEARch:PROTocol:ARINc:CONDition is set to LDATa.
Parameters:
<DataMaximum> 01X-string