Remote control commands
R&S
®
RTM3000
451User Manual 1335.9090.02 ─ 09
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<State> ON | OFF
*RST: OFF
PROBe<m>:SETup:DCOFfset?
PROBe<m>:PMETer:RESults:SINGle?
rcset
Suffix:
<m>
.
1..4
Return values:
<ResultValue>
Usage: Query only
PROBe<m>:PMETer:RESults:COMMon?
The command is available only for Rohde & Schwarz differential and modular probes
that have an R&S ProbeMeter.
It returns the common mode voltage, which is the mean voltage between the signal
sockets and the ground socket.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<ResultValue> Default unit: V
Usage: Query only
PROBe<m>:PMETer:RESults:DIFFerential?
The command is available only for Rohde & Schwarz differential and modular probes
that have an R&S ProbeMeter.
It returns the differential voltage, which is the voltage between the positive and nega-
tive signal sockets.
Suffix:
<m>
.
1..4
Return values:
<ResultValue>
Usage: Query only