Remote control commands
R&S
®
RTM3000
524User Manual 1335.9090.02 ─ 09
Usage: Query only
MEASurement<m>:STATistics:VALue:ALL?
Returns all values from the statistics buffer.
Note: Valid buffered values can only be read if the acquisition is stopped. As long as
the acquisition is running, the buffer contents is changing and the buffered values are
not valid for reading.
Suffix:
<m>
.
1..8
Selects the measurement place.
Return values:
<ValueList> Comma separated List of Values
Comma-separated list of statistical values
Usage: Query only
MEASurement<m>:STATistics:VALue<n>?
Returns the n
th
statistical value from the indicated buffer place.
Note: Valid buffered values can only be read if the acquisition is stopped. As long as
the acquisition is running, the buffer contents is changing and the buffered values are
not valid for reading.
Suffix:
<m>
.
1..8
Selects the measurement place.
<n> *
Buffer place. The buffer size is 1000 results.
Return values:
<StatisticValue> Statistical value
Usage: Query only
EXPort:MEASurement<m>:STATistics:NAME
Defines the path and filename of the statistics file. The file format is CSV. If the file
already exists, it will be overwritten without notice.
Suffix:
<m>
.
1..8
Selects the measurement place.
Parameters:
<FileName> String parameter
Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals