Remote control commands
R&S
®
RTM3000
580User Manual 1335.9090.02 ─ 09
DIGital<m>:DATA:YORigin?
REFCurve<m>:DATA:YORigin?
Return the voltage value for binary value 0 of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(FORM UINT, 8|16|32).
Suffix:
<m>
.
1..4
Return values:
<Yorigin> Voltage in V
Example: See Chapter 17.2.1.4, "Reading waveform data in unsigned inte-
ger format", on page 426
Usage: Query only
CHANnel<m>:DATA:YINCrement?
CHANnel<m>:DATA:ENVelope:YINCrement?
CALCulate:MATH<m>:DATA:YINCrement?
MASK:DATA:YINCrement?
LOGic<p>:DATA:YINCrement?
DIGital<m>:DATA:YINCrement?
REFCurve<m>:DATA:YINCrement?
Return the voltage value per bit of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(FORM UINT, 8|16|32).
Suffix:
<m>
.
1..4
Return values:
<Yincrement> Voltage in V
Example: See Chapter 17.2.1.4, "Reading waveform data in unsigned inte-
ger format", on page 426
Usage: Query only
CHANnel<m>:DATA:YRESolution?
CHANnel<m>:DATA:ENVelope:YRESolution?
CALCulate:MATH<m>:DATA:YRESolution?
MASK:DATA:YRESolution?
LOGic<p>:DATA:YRESolution?
DIGital<m>:DATA:YRESolution?
REFCurve<m>:DATA:YRESolution?
Return the vertical bit resolution of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(FORM UINT, 8|16|32).