EasyManua.ls Logo

Rohde & Schwarz RTM3K-COM2

Rohde & Schwarz RTM3K-COM2
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTM3000
609User Manual 1335.9090.02 ─ 09
Parameters:
<MisoSource> CH1 | CH2 | CH3 | CH4 | NONE | D0..D15
*RST: NONE
BUS<b>:SPI:DATA:POLarity <Polarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1) on the
data line.
Suffix:
<b>
.
1..4
Parameters:
<Polarity> POSitive | NEGative
POSitive = high active
NEGative = low active
*RST: POSitive
BUS<b>:SPI:MOSI:POLarity
<MosiPolarity>
Selects if transmitted data is high active (high = 1) or low active (low = 1) on the MOSI/
MISO line.
Suffix:
<b>
.
1..4
Parameters:
<MosiPolarity> ACTLow | ACTHigh
*RST: ACTH
BUS<b>:SPI:MISO:POLarity <MisoPolarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1) on the
MISO line.
Suffix:
<b>
.
1 | 3. Bus 2 or 4 is not available if the MISO source is used on
bus 1 or 3, respectively.
Parameters:
<MisoPolarity> ACTLow | ACTHigh
*RST: ACTH
BUS<b>:SPI:BORDer <BitOrder>
Defines if the data of the messages starts with MSB (most significant bit) or LSB (least
significant bit).
Suffix:
<b>
.
1..4
Serial bus analysis

Table of Contents

Related product manuals