R3765/67G Series Network Analyzer Operation Manual
8.2 Error Messages
8-8
Operating Error
Register save error.
An error occurred in saving a register.
<How to handle>
Available space is not in C: drive.
Delete unnecessary files. (Refer to section 7.14.)
Segment #x error.
The PROGRAM SWEEP or USER SWEEP was specified in a
state that STOP FREQ of the Xth segment is higher than START
FREQ of the following segment.
<How to handle>
Specify the frequency of the Xth segment again. (Refer to sub-
section 7.9.2 and 7.9.3.)
Segment not entered.
The PROGRAM SWEEP or USER SWEEP was specified without
setting any segment.
<How to handle>
Specify the segment. (Refer to sub-section 7.9.2 and 7.9.3.)
Some STD not
memorized.
To execute the DONE operation was attempted without obtaining
all related calibration data.
<How to handle>
Obtain all calibration data (OPEN, SHORT, LOAD). (Refer to
sub-section 7.5.5.)