R3765/67G Series Network Analyzer Operation Manual
8.2 Error Messages
8-7
Operating Error
Memory not found.
A trace operation (DATA/MEM, etc.) or a memory trace display
(DISPLAY MEMORY, DISPLAY DATA&MEM) was specified,
with a memory trace not stored.
<How to handle>
Obtain the memory trace. (Refer to sub-section 7.4.4 and 7.4.7.)
None Controller
A plot output was specified not in system controller mode.
<How to handle>
Set to the system controller mode. (Refer to sub-section 7.13.2.)
Outside calibration
range
Outside the calibration range
The sweep range is set outside the calibration range.
<How to handle>
The calibration data cannot be used because the setting
conditions are different from those used when the calibration data
was acquired. Set the sweep range to a range where the sweep
can be calibrated, or obtain the calibration data again (refer to
Section 7.5).
Please set 1-trace
FORMAT.
With the measurement format two traces (LOGMAG&PHASE,
LOGMAG&DELAY, LOGMAG&PHASE), the memory trace
display (DISPLAY MEMORY, DISPLAY DATA&MEM) was
specified.
<How to handle>
The memory trace display is invalid with the measurement format
two traces.
Set the measurement format to one trace (other than LOGMAG&
PHASE, LOGMAG&DELAY or LOGMAG&PHASE). (Refer to
sub-section 7.4.2 and 7.4.4.)
Register recall error.
An error occurred in recalling a register.
<How to handle>
An unsaved register was specified.
The register was damaged.
Clear the register with CLEAR REG and save it again. (Refer to
sub-section 7.14.1.)