R3765/67G Series Network Analyzer Operation Manual
A.3 Initial Setting
A-7
* 2 : Start/Stop level
* 3: Selection item of input port
* 4: Measurement format
Type
Power on or preset *RST
Start Stop Start Stop
AG -3dBm 0dBm -3dBm 17dBm
AG+OPT10 -8dBm 0dBm -8dBm 12dBm
BG -13dBm 0dBm -13dBm 7dBm
BG+OPT10 -18dBm 0dBm -18dBm 2dBm
BG+12 -11dBm 0dBm -11dBm 9dBm
BG+OPT10+OPT12 -16dBm 0dBm -16dBm 4dBm
CG, CG+OPT11/14 -10dBm 0dBm -10dBm 10dBm
CG+OPT12/13 -16dBm 0dBm -16dBm 4dBm
CG+OPT10,
CG+OPT10+OPT11/14
-15dBm 0dBm -15dBm 5dBm
CG+OPT10+OPT12/13 -21dBm -1dBm -21dBm -1dBm
Type Channel CH1 CH2 CH3 CH4
AG A/R B/R A/R B/R
BG
REFLECTION
TRANSMISSION
REFLECTION
TRANSMISSION
CG S11S21S11S21
Type Channel CH1 CH2 CH3 CH4
AG LOG MAG LOG MAG LOG MAG LOG MAG
BG LOG MAG LOG MAG POLAR LOG MAG
CG LOG MAG LOG MAG POLAR LOG MAG