CALIBRATING AV7000: DYNAMIC CALIBRATION
PRODUCT REFERENCE GUIDE
353
6. Adjust the gain until the white level shown in the Image Viewer is between 180
and 200.
NOTE: If pulsed illumination is enabled in Device Settings | Camera n |
Imaging Global Gain with Pulsed Illumination should be adjusted. If pulsed
illumination is not enabled, make adjustments to Global Gain with Continu-
ous Illumination.