Remote Control Commands
R&S
®
SMW200A
1077User Manual 1175.6632.02 ─ 30
Example: See "Programming Examples" on page 1071
Manual operation: See "Carrier Spacing" on page 401
[:SOURce<hw>]:BB:MCCW:CLOCk?
Queries the output clock rate. The output clock rate depends on the number of carriers
and on the selected carrier spacing.
Return values:
<Clock> float
Range: 0 to Max
Increment: 1E-3
*RST: 0
Example: See "Programming Examples" on page 1071
Usage: Query only
Manual operation: See "Clock Frequency" on page 401
[:SOURce<hw>]:BB:MCCW:CFACtor:MODE <Mode>
Sets the mode by which automatic settings minimize the crest factor or hold it at a
chosen value.
Parameters:
<Mode> OFF | CHIRp | SLOW
SLOW
corresponds to the manual control "Target Crest"
*RST: CHIRp
Example: See "Programming Examples" on page 1071
Manual operation: See "Optimize Crest Factor Mode" on page 402
[:SOURce<hw>]:BB:MCCW:CFACtor <CFactor>
Sets the desired crest factor, if the optimization mode target crest factor is used.
Parameters:
<CFactor> float
Range: 0 to 30
Increment: 0.01
*RST: 3
Example: See "Programming Examples" on page 1071
Manual operation: See "Desired Crest Factor" on page 402
SOURce Subsystem