Remote Control Commands
R&S
®
SMW200A
1079User Manual 1175.6632.02 ─ 30
Return values:
<State[,State..]> select
Example: See "Programming Examples" on page 1071
Manual operation: See "Carrier Table" on page 404
[:SOURce<hw>]:BB:MCCW:CARRier:PHASe <CarrierIndex>, <Phase>
For disabled optimization of the crest factor, sets the start phase of the selected carrier.
Parameters:
<CarrierIndex> integer
Range: 0 to lastCarrier
<Phase> float
Range: 0 to 359.99
Increment: 0.01
*RST: 0
Default unit: DEG
Example: see "Programming Examples" on page 1071
Manual operation: See "Carrier Table" on page 404
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:PHASe <Phas0[,Phas1..]>]
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:PHASe? [<Start>[, <Count>]]
Sets the start phase of the carrier with the aid of a value list.
Setting parameters:
<Phas0[,Phas1..]> float
Range: 0 to 360
Increment: 0.01
*RST: 0
Default unit: DEG
Query parameters:
<Start> integer
start carrier index
Range: 0 to lastCarrier
<Count> integer
number of carriers in the carrier range, starting from the
<Start> carrier
Range: 1 to max
Return values:
<Phas[,Phas..]> float
Example: see "Programming Examples" on page 1071
Manual operation: See "Carrier Table" on page 404
SOURce Subsystem