Remote Control Commands
R&S
®
SMW200A
1155User Manual 1175.6632.02 ─ 30
Manual operation: See "Bandwidth" on page 556
[:SOURce<hw>]:LFOutput<ch>:FREQuency <Frequency>
Sets the frequency of the LF signal in [:SOURce<hw>]:LFOutput:FREQuency:
MODE CW|FIXed mode.
Note:
●
If the LF generator is used as a signal source, the instrument performs the analog
modulations (AM/FM/ΦM/PM) with this frequency.
●
In sweep mode ([:SOURce<hw>]:LFOutput:FREQuency:MODE SWE), the fre-
quency is coupled with the sweep frequency.
Parameters:
<Frequency> float
Range: 0.1 to depends on the installed options
(R&S SMW-K24)
Increment: 0.01
*RST: 1000
Default unit: Hz
Example: See Example "Configuring the LF generator" on page 1153.
Manual operation: See "Frequency" on page 554
[:SOURce<hw>]:LFOutput<ch>:PERiod?
Queries the repetition frequency of the sine signal.
Return values:
<LfSinePeriod> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.001
Default unit: s
Example: See Example "Configuring the LF generator" on page 1153.
Usage: Query only
[:SOURce<hw>]:LFOutput:FREQuency:MANual <Manual>
Sets the frequency of the subsequent sweep step if LFO:SWE:MODE MAN.
Use a separate command for each sweep step.
SOURce Subsystem