EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Page 1170

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1170User Manual 1175.6632.02 ─ 30
Parameters:
<Dwell> <Dwell#1>{, <Dwell#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data. The list of numbers can be of any length, with the list
entries separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy. See also :
FORMat[:DATA] on page 879 for more details.
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Manual operation: See "Edit List Mode Data" on page 525
[:SOURce<hw>]:LIST:DWELl:LIST:POINts?
Queries the number (points) of dwell time entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Usage: Query only
[:SOURce<hw>]:LIST:FREQuency <Frequency>
Enters the frequency values in the selected list.
Parameters:
<Frequency> <Frequency#1>{, <Frequency#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data.
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also :FORMat[:DATA].
Range: 300 kHz to RFmax (depends on the installed
options)
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Manual operation: See "Edit List Mode Data" on page 525
SOURce Subsystem

Table of Contents