1171User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:LIST:FREQuency:POINts?
Queries the number (points) of frequency entries in the seleced list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Usage: Query only
[:SOURce<hw>]:LIST:INDex <Index>
Sets the list index in LIST:MODE STEP.
After the trigger signal, the instrument processes the frequency and level settings of
the selected index.
Parameters:
<Index> integer
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Manual operation: See "Current Index" on page 519
[:SOURce<hw>]:LIST:INDex:STARt <Start>
[:SOURce<hw>]:LIST:INDex:STOP <Stop>
Sets the start and stop index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list.
Parameters:
<Stop> integer
Index range
Only values inside this range are processed in list mode
Range: 0 to list length
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 1165.
Manual operation: See "List Range from/to" on page 521
SOURce Subsystem