Remote Control Commands
R&S
®
SMW200A
1223User Manual 1175.6632.02 ─ 30
:SYSTem:ERRor:STATic?
Returns a list of all errors existing at the time when the query is started. This list corre-
sponds to the display on the info page under manual control.
Return values:
<StaticErrors> string
Example: See Example "Querying the error queue" on page 1218
Usage: Query only
Manual operation: See "Static Errors/Error History" on page 1261
:SYSTem:DLOCk
<DispLockStat>
Disables the manual operation via the display, including the front panel keyboard of the
instrument and the LOCAL key.
Parameters:
<DispLockStat> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
SYST:DLOC ON
Activates the display lock. The instrument cannot be operated
via the display until it has been enabled with SYST:DLOC OFF.
Manual operation: See "User Interface" on page 758
See "Enabling a locked user interface for manual operation"
on page 760
:SYSTem:KLOCk <State>
Disables the front panel keyboard of the instrument including the LOCAL key.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
SYST:KLOC ON
Locks the front panel and external controls.
To enable the controls, send SYST:KLOC OFF.
Manual operation: See "User Interface" on page 758
See "Enabling a locked user interface for manual operation"
on page 760
:SYSTem:NINFormation?
Queries the oldest information message ("Error History > Level > Info") in the error/
event queue.
SYSTem Subsystem