EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Page 1235

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1235User Manual 1175.6632.02 ─ 30
:SYSTem:SECurity:SANitize[:STATe] <SecPassWord>, <MmemProtState>
Sanitizes the internal memory.
Parameters:
<MmemProtState> 0 | 1 | OFF | ON
*RST: 0
Setting parameters:
<SecPassWord> string
Example:
SYSTem:SECurity:SANitize[:STATe] 1
Manual operation: See "Sanitize" on page 757
:SYSTem:SECurity:SUPolicy
<SecPassWord>, <UpdatePolicy>
Configures the automatic signature verification for firmware installation.
Parameters:
<UpdatePolicy> STRict | CONFirm | IGNore
*RST: n.a. (factory preset: CONFirm)
Setting parameters:
<SecPassWord> string
Manual operation: See "Secure Update Policy" on page 755
:SYSTem:SPECification? <Id>
Retrieves data sheet information for a specific parameter.
Setting parameters:
<Id> string
Identifies the name of the entry in the data sheet, as quieried
with the command :SYSTem:SPECification:
IDENtification:CATalog? on page 1236
Return values:
<ValList> float
Comma-separated list with the specified and, if available, the
typical value of the parameter, as specified in the data sheet.
SYSTem Subsystem

Table of Contents