Maintenance, Storage and Disposal
R&S
®
SMW200A
1276User Manual 1175.6632.02 ─ 30
Calibrating the touchscreen
The screen responds to the touch of your finger, that means you can select or activate
an item by tapping it. Due to the design of the touchscreen, the capture area is firmly
set to a certain size and you do not need to define a range. If the instrument does not
respond correctly to the touch, we recommend that you calibrate the touchscreen.
1. Press the [SETUP] key.
2. Select the "Maintenance > Check Front Panel"
The "Check Front Panel" opens.
3. Select "Calibrate".
4. Tap the calibration points.
The instrument completes the calibration process automatically and returns to the
"Check Front Panel" dialog.
Testing the key panel
To perform the key panel test, you operate the keys at the front panel, and check the
response of the instrument in the "Check Front Panel" dialog. To perform this test prop-
erly, check each key of the front panel. The test is only completed, when you have veri-
fied all keys.
During the test, the actual functions of the keys are disabled.
1. Press the [SETUP] key.
2. Select "Maintenance > Check Front Panel"
The "Check Front Panel" dialog opens.
Performing Maintenance Tasks