Signal Routing and System Configuration
R&S
®
SMW200A
203User Manual 1175.6632.02 ─ 30
14. In the block diagram, select "RF > On".
The R&S SMW performs the following:
Retrieves the R&S SZU calibration data.
Indicates the connected R&S SZU in the block diagram.
Sets the output signal level and frequency to the default values of the
R&S SZU and indicates the value in the status bar.
1 2 3 4 5
6
7
9
10
8
Figure 4-27: Block diagram of wideband baseband R&S
SMW (R&S
SMW-B9) with connected R&S
SZU
1 = RF frequency at the R&S SZU RF output
2, 5 = RF frequency and level settings at the second R&S SMW RF output
3 = I/Q modulator of the R&S SZU
4 = RF level, nominal level at the R&S SZU RF output (regulated level)
6 = R&S SZU, indicated as waveguide port icon, label and the serial number
6, 7 = Access to the R&S SZU RF settings (RF block or icon)
8 = Active I/Q analog output, firmly routed to the R&S SZU; the example uses differential output signal
9 = I/Q analog B is disabled, because the differential output of the first path is used
10 = R&S SMW RF settings
To create a remote control sequence or modify the initialization sequence
You have connected and configured external instruments to the R&S SMW as descri-
bed in "To connect instruments to the I/Q analog interface" on page 196.
How to Connect External Instruments and Configure the Signal Flow