Signal Routing and System Configuration
R&S
®
SMW200A
208User Manual 1175.6632.02 ─ 30
See "To start signal generation" on page 210
To cable the instruments
Refer to Figure 4-29 for illustration the required connections. The test setup does not
show the LAN connections.
1. Connect all six instruments to a LAN.
2. Use suitable connecting cables.
The connecting cables must have the same length and type, applies for the follow-
ing connections:
REF OUT to REF IN
BBMM1/2 to DIG IQ
USER6 to both USER3 connectors
RF outputs to DUT
3. Avoid unnecessary cable lengths and branching points.
4. Provide the instruments with common reference frequency signal.
In this test setup, the internal reference frequency of R&S SMW#1 is distributed to
the other instruments. The instruments are connected in a daisy chain.
5. Use a T-connector to distribute the trigger signal generated by R&S SMW#1 and to
trigger both instruments simultaneously:
a) Connect the instruments in a star network, concerning the trigger signal
b) Connect the USER 6 output of the R&S SMW#1 to the USER 3 input of
R&S SMW#2 and the R&S SMW#1 itself.
In this test setup, the instruments are triggered by the trigger signal of the first one.
For more information, see Chapter 10.3.2, "Triggering Several Instruments with a
Common Trigger Signal", on page 648.
6. Connect the external instruments to the digital I and Q connectors (BBMM 1 and 2)
of the R&S SMW.
Tip: See also "To find out the correct connector" on page 190.
To configure the R&S SMW that simulates the first subset
1. Select "System Configuration" and configure:
a) A 1x8x8 MIMO scenario with coupled sources.
See "To simulate complex MxN MIMO scenarios using the "Fading and Base-
band Configuration" settings" on page 191.
b) "Subset = 1"
c) Map the I/Q streams to the output interfaces as follows: Stream A and B to RF
A and B; stream C and D to BBMM 1/2 respectively.
See Chapter 4.9.4, "How to Route the I/Q Streams to the Output Interfaces",
on page 193
2. Set "System Config. > Setup > Global Connectors" > "User 6 > Direction = Out-
put".
How to Connect External Instruments and Configure the Signal Flow