Configuring the RF Signal
R&S
®
SMW200A
494User Manual 1175.6632.02 ─ 30
Auto mode (Sweep/List)
Figure 8-9: Auto mode (Sweep/List)
The instrument generates the signal continuously.
Trigger mode "Auto" is prerequisite (default). It causes the continuous generation
of the sweep signal.
Starts signal generation immediately with "State = On".
Switches automatically to the next step when the Dwell time has elapsed.
Stops signal generation with "State = Off".
Table 8-2: Cross-reference between manual and remote control in Auto mode (Sweep/List)
Manual control mode:
"Auto"
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:TRIGger<hw>:FSWeep:SOURce AUTO [IMMediate]
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep
:TRIGger<hw>:PSWeep:SOURce AUTO [IMMediate]
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep
:TRIGger<hw>:LFFSweep:SOURce AUTO [IMMediate]
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO
List
:SOURce<hw>:LIST:MODE AUTO
:SOURce<hw>:LIST:TRIGger:SOURce AUTO
Varying the RF Signal in List or Sweep Mode