Configuring the RF Signal
R&S
®
SMW200A
495User Manual 1175.6632.02 ─ 30
Single / Extern Single mode (Sweep/List)
Figure 8-10: Single / Extern Single mode (sweep / list )
In single mode, you can specify, that the signal returns to the start value when a sweep
cycle has been completed.
See "The Retrace function" on page 500 for details.
The instrument generates a single sweep cycle.
Trigger mode "Manual". A trigger event initiates one sweep from the start value to
the end value.
"State = On" sets the signal to the start value: the sweep start frequency, the
sweep start power or the frequency-power value pair of the selected index in the
list.
Starts signal generation with a trigger event.
Switches automatically to the next step when the Dwell time has elapsed.
Stops signal generation at the set end value and waits for the subsequent trigger
event.
Trigger sources:
The corresponding remote control command.
An externally applied trigger signal.
"State = Off" stops the signal generation in sweep or list mode.
Varying the RF Signal in List or Sweep Mode