Multi-Instrument Setups
R&S
®
SMW200A
666User Manual 1175.6632.02 ─ 30
3. Select "RF Ports / Instruments > RF ports table" > "Additional S-Parameter
Files".
In this dialog, you can load S-parameter files to compensate for exchanged cables
or other components.
Settings:
Add S-Parameter File..................................................................................................666
S-Parameter Filename................................................................................................666
In/Out.......................................................................................................................... 666
Delete..........................................................................................................................666
Add S-Parameter File
Appends a row in the table with S-parameter files.
S-Parameter Filename
Indicates the filename of the loaded file.
Select the filed to open the standard "File Select" dialog for selecting and loading S-
parameters (Touchstone) files. S-parameters files are files with predefined extension
(*.s<n>p) and file format.
Use the standard "File Manager" function to transfer external data lists to the instru-
ment.
In/Out
Sets the origin and destination ports, i.e. the port number from that the signal is coming
and the port to that it is going. Use this parameter to change the orientation of the sig-
nal chain.
Delete
Removes the selected table row and the S-parameter file.
Generating Phase-Coherent Synchronous and Aligned Signals