Getting Started
R&S
®
SMW200A
72User Manual 1175.6632.02 ─ 30
3. Tap the "Save" button.
The file MyTestSignal.savrcltxt is saved in the default directory /var/user.
To load saved instrument settings
You can restore the settings to the instrument at any time using the settings file.
1. Press the [PRESET] button to restore the default instrument settings so you can
check that the saved user settings are restored afterwards.
2. Press the [SAVE/RCL] key.
3. In the "Save/Recall" dialog, select "Recall" operation.
4. Navigate to the directory, in which you saved the file MyTestSignal and select it.
5. Tap the "Recall" button.
All instrument settings are restored and the display resembles Figure 3-16, which
shows the instrument display right before the settings were saved.
How to display all parameters with values different to their preset values
When you load a file to your instrument, you may do not have enough information on
the changed settings. In such case, it is useful to visualize all parameters that have
been changed from their default state.
Trying Out the Instrument