EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Page 899

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
899User Manual 1175.6632.02 ─ 30
Manual operation: See "Filter Length" on page 592
:SENSe<ch>[:POWer]:FILTer:NSRatio <NSRatio>
Sets an upper limit for the relative noise content in fixed noise filter mode (:
SENSe<ch>[:POWer]:FILTer:TYPE). This value determines the proportion of intrin-
sic noise in the measurement results.
Parameters:
<NSRatio> float
Range: 0.001 to 1
Increment: 0.001
*RST: 0.01
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 891.
Manual operation: See "Noise/Signal Ratio" on page 592
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
<MTime>
Sets an upper limit for the settling time of the auto-averaging filter in the NSRatio
mode and thus limits the length of the filter. The filter type is set with command :
SENSe<ch>[:POWer]:FILTer:TYPE.
Parameters:
<MTime> float
Range: 1 to 999.99
Increment: 0.01
*RST: 4
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 891.
Manual operation: See "Timeout" on page 592
:SENSe<ch>[:POWer]:FILTer:SONCe
Starts searching the optimum filter length for the current measurement conditions. You
can check the result with command :SENS1:POW:FILT:LENG:USER? in filter mode
USER (:SENSe<ch>[:POWer]:FILTer:TYPE).
Example:
SENS1:FILT:TYPE USER
Selects user filter mode.
:SENS1:FILT:SONC
Activates the search for the optimum filter length.
:SENS1:FILT:LENG?
Returns the found optimum filter length.
Response: 128
Usage: Event
SENSe, READ, INITiate and SLISt Subsystems

Table of Contents