EasyManua.ls Logo

Rohde & Schwarz RTM3K-COM2

Rohde & Schwarz RTM3K-COM2
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTM3000
437User Manual 1335.9090.02 ─ 09
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Range> Voltage range value
Range: 8e-3 to 80 (without probe attenuation)
*RST: 40e-3
Default unit: V
CHANnel<m>:POSition <Position>
Sets the vertical position of the waveform in divisions. While the offset sets a voltage,
position is a graphical setting given in divisions.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Position> Graphical position
Range: -5 to 5
*RST: 0
Default unit: div
CHANnel<m>:OFFSet <Offset>
Sets the offset voltage, which is subtracted to correct an offset-affected signal.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Offset> Offset value
Range: Depend on vertical scale and probe attenuation.
Increment: Depends on vertical scale and probe attenuation.
*RST: 0
Default unit: V
CHANnel<m>:COUPling <Coupling>
Selects the connection of the indicated channel signal - coupling and termination.
Waveform setup

Table of Contents

Related product manuals