Remote control commands
R&S
®
RTM3000
449User Manual 1335.9090.02 ─ 09
PROBe<m>:SETup:ZAXV <ExtAttZA15>
This command is relevant only for differential active probes R&S RT-ZD10, R&S RT-
ZD20 or R&S RT-ZD30. If you use the external attenuator R&S RT-ZA15, enable it to
include the external attenuation in the measurements.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ExtAttZA15> ON | OFF
*RST: OFF
PROBe<m>:SETup:ADVanced:AUDioverload <ZHDAudioState>
This command is relevant for R&S RT-ZHD probes. It activates the acoustic overrange
warning in the probe control box.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ZHDAudioState> ON | OFF
*RST: OFF
PROBe<m>:SETup:ADVanced:FILTer <ZHDLowpassState>
This command is relevant for R&S RT-ZHD probes. It activates the lowpass filter in the
probe control box.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ZHDLowpassState> ON | OFF
*RST: OFF
PROBe<m>:SETup:ADVanced:RANGe <ZHDRange>
This command is relevant for R&S RT-ZHD probes. It sets the voltage range of the
probe.
Waveform setup