Remote control commands
R&S
®
RTM3000
458User Manual 1335.9090.02 ─ 09
0
The number of acquired waveforms is less than the number
required for average calculation. See ACQuire:AVERage:
COUNt on page 457.
1
The instrument acquired a sufficient number of waveforms to
determine the average.
Usage: Query only
TIMebase:ROLL:AUTomatic
<AutomaticRoll>
Enables the automatic roll mode. The instrument switches to roll mode if the timebase
is equal or slower than the roll mode limit defined with TIMebase:ROLL:MTIMe.
Parameters:
<AutomaticRoll> ON | OFF
*RST: OFF
TIMebase:ROLL:MTIMe <MinTimeBase>
The roll mode is enabled automatically if the time base exceeds the MinTimeBase,
and if TIMebase:ROLL:AUTomatic on page 458 is set ON.
Parameters:
<MinTimeBase> Limit value for roll mode enabling.
*RST: 500e-3
Default unit: s/div
ACQuire:INTerpolate <InterpolationType>
Defines the interpolation mode.
See also: "Interpolation" on page 75
Parameters:
<InterpolationType> SINX | LINear | SMHD
LINear
Linear interpolation between two adjacent sample points.
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample & hold causes a histogram-like interpolation.
*RST: SINX
Waveform setup

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals