Remote control commands
R&S
®
RTM3000
519User Manual 1335.9090.02 ─ 09
Suffix:
<m>
.
1..8
Selects the measurement place.
Parameters:
<SignalSlope> POSitive | NEGative
Slope of first waveform ("Measure Source").
*RST: POS
<ReferenceSlope> POSitive | NEGative
Slope of the second waveform ("Measure Source 2"). The value
is ignored in delay to trigger measurements.
*RST: POS
MEASurement<m>:DELay:DIRection <Direction>,<Direction>
Selects the edges for delay measurement and delay to trigger measurement.
See also: Chapter 8.2.4, "Delay setup", on page 157.
Suffix:
<m>
.
1..8
Selects the measurement place.
Parameters:
<Direction> NEARest | FRFI | FRLA
For delay measurement, it defines the edge of the first wave-
form ("Measure Source").
NEARest: Nearest edge to the trigger point.
FRFI: First edge from the left window edge.
FRLA: First edge from the right window edge, which is the last
edge if counted from the left window edge.
For delay to trigger measurement, it defines the measured
edge.
NEARest: Nearest edge to the trigger point.
FRFI: Next edge from the trigger to the right.
FRLA: Next edge from the trigger to the left.
<Direction>
NEARest | FRFI | FRLA
For delay measurement, it defines the edge of the second wave-
form ("Measure Source 2").
NEARest: Nearest edge to the edge determined on "Measure
Source".
FRFI: Next edge to the right, counted from the edge determined
on "Measure Source".
FRLA: Next edge to the left, counted from the edge determined
on "Measure Source"
For delay to trigger measurement, enter any value. This value is
ignored.
Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals