EasyManua.ls Logo

Rohde & Schwarz RTM3K-COM2

Rohde & Schwarz RTM3K-COM2
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTM3000
553User Manual 1335.9090.02 ─ 09
SPECtrum:MARKer:REFerence:SETup:SPAN
Sets the span range, the ratio of the capture range and the width of the specified refer-
ence mode, if SPECtrum:MARKer:REFerence:SETup:MODE is set to RANGe.
Parameters:
<ReferenceSpan>
SPECtrum:MARKer:RMARker?
Returns the frequency and the level values of the present reference marker.
Return values:
<ReferenceFrequency>
<ReferenceLevel>
Usage: Query only
SPECtrum:MARKer:RMARker:FREQuency?
Returns the frequency of the reference marker.
Return values:
<ReferenceFrequency>
Usage: Query only
SPECtrum:MARKer:RMARker:LEVel?
Returns the level of the reference marker.
Return values:
<ReferenceLevel>
Usage: Query only
SPECtrum:MARKer:RMODe <ResultMode>
Defines the values that are shown in the peak list. You can use absolute values, or
delta values in comparison to the reference marker.
Parameters:
<ResultMode> ABSolute | FREQuency | LEVel | FLEVel
ABSolute
Frequency and magnitude in absolute values.
FREQuency
Frequency relative to the reference marker frequency, magni-
tude in absolute values.
LEVel
Frequency in absolute values, magnitude relative to the level of
the reference marker.
Applications

Table of Contents