Remote control commands
R&S
®
RTM3000
642User Manual 1335.9090.02 ─ 09
Parameters:
<IdentifierType> B11 | B29 | ANY
ANY: use if the identifier length is not relevant. Not available for
trigger type ID.
*RST: B11
TRIGger:A:CAN:ICONdition <IdentifierCondition>
Sets the comparison condition: If the pattern contains at least one X (don't care), you
can trigger on values equal or not equal to the specified value. If the pattern contains
only 0 and 1, you can also trigger on a range greater than or lower than the specified
value.
The command is relevant, if TRIGger:A:CAN:TYPE is set to ID or IDDT.
Parameters:
<IdentifierCondition> EQUual | NEQual | GTHan | LTHan
*RST: EQ
TRIGger:A:CAN:IDENtifier <Identifier>
Defines the identifier pattern. The pattern length is defined with TRIGger:A:CAN:
ITYPe.
The command is relevant if TRIGger:A:CAN:TYPE is set to ID or IDDT.
Parameters:
<Identifier> String containing binary pattern with 11 bit or 29 bit. Characters
0, 1, and X are allowed.
TRIGger:A:CAN:DCONdition <DataCondition>
Sets the comparison condition for data: If the pattern contains at least one X (don't
care), you can trigger on values equal or not equal to the specified value. If the pattern
contains only 0 and 1, you can also trigger on a range greater than or lower than the
specified value.
The command is relevant if TRIGger:A:CAN:TYPE is set to IDDT.
Parameters:
<DataCondition> EQUal | NEQual | GTHan | LTHan
*RST: EQ
TRIGger:A:CAN:DLC <DataLength>
Defines the length of the data pattern - the number of bytes in the pattern.
The command is relevant if TRIGger:A:CAN:TYPE is set to IDDT.