Remote control commands
R&S
®
RTM3000
776User Manual 1335.9090.02 ─ 09
17.13.2.2 Parallel clocked bus configuration
BUS<b>:CPARallel:CLOCk:SOURce............................................................................... 776
BUS<b>:CPARallel:CLOCK:SLOPe................................................................................. 776
BUS<b>:CPARallel:CS:ENABle.......................................................................................776
BUS<b>:CPARallel:CS:SOURce......................................................................................776
BUS<b>:CPARallel:CS:POLarity......................................................................................777
BUS<b>:CPARallel:CLOCk:SOURce <ClockSource>
Selects the logic channel that is used as clock line.
Suffix:
<b>
.
1..2
Selects the parallel bus.
Parameters:
<ClockSource> D0..D15
*RST: D0
BUS<b>:CPARallel:CLOCK:SLOPe <ClockSlope>
Selects if the data is sampled on the rising or falling slope of the clock, or on both
edges (EITHer). The clock slope marks the begin of a new bit.
Suffix:
<b>
.
1..2
Selects the parallel bus.
Parameters:
<ClockSlope> POSitive | NEGative | EITHer
BUS<b>:CPARallel:CS:ENABle <ChipSelectEnable>
Enables and disables the chip select line.
Suffix:
<b>
.
1..2
Selects the parallel bus.
Parameters:
<ChipSelectEnable> ON | OFF
*RST: ON
BUS<b>:CPARallel:CS:SOURce <ChipSelectSource>
Selects the logic channel that is used as chip select line.
Suffix:
<b>
.
1..2
Selects the parallel bus.
Mixed signal option (option R&S RTM-B1)