Remote control commands
R&S
®
RTM3000
780User Manual 1335.9090.02 ─ 09
WGENerator:VOLTage:OFFSet <Offset>
Sets the DC offset of the selected generator function.
Parameters:
<Offset> Numeric value
Range: -3.00000E+00 to 3.00000E+00
*RST: 0.00E+00
Default unit: V
WGENerator:FREQuency <Frequency>
Defines the frequency.
Parameters:
<Frequency> Range: Depends on the selected function. For details, refer
to the datasheet.
*RST: 10 kHz
WGENerator:FUNCtion:PULSe:DCYCle <DutyCycle>
Defines the duty cycle value of the generator function pulse.
Parameters:
<DutyCycle> Numeric value
Range: 1.000E+01 to 9.000E+01
*RST: 2.500E+01
Default unit: %
WGENerator:TRIangle:SYMMetry <Symmetry>
Sets the symmetry for the triangle waveform.
Parameters:
<Symmetry> 50% defines symmetric triangles. Values <50% define triangles
with steeper rising edge leaned to the left. Values >50% define
triangles with steeper falling edge leaned to the right.
Range: 1 to 99
Increment: 1
*RST: 50
Default unit: %
WGENerator:FUNCtion:PULSe:ETIMe <EdgeTime>
Sets the pulse edge time.
Parameters:
<EdgeTime>
Signal generation (option R&S RTM-B6)