R3765/67G Series Network Analyzer Operation Manual
11 SPECIFICATIONS
11-3
Accuracy R3765CG/67CG: ±0.5dB (50MHz, 0dBm, 23°C ± 5°C)
For OPT 10, OPT 10+OPT 11 or OPT 10+OPT 14
±0.5dB (50MHz, -5dBm, 23°C ± 5°C)
For OPT 12 or OPT 13
±0.5dB (50MHz, -6dBm, 23°C ± 5°C)
For OPT 10+OPT 12 or OPT 10+OPT 13
±0.5dB (50MHz, -11dBm, 23°C ± 5°C)
For OPT 11, 13 or 14, the accuracy is
specified using TEST PORT 1.
If OPT 10 is installed, the accuracy is
specified using an attenuation of 0 dB.
R3765BG/67BG: ±0.5dB (50MHz, -3dBm, 23°C ± 5°C)
For OPT 10 ±0.5dB (50MHz, -8dBm, 23°C ± 5°C)
For OPT 12 ±0.5dB (50MHz, -1dBm, 23°C ± 5°C)
For OPT 10+OPT 12 ±0.5dB (50MHz, -6dBm, 23°C ± 5°C)
If OPT 10 is installed, the accuracy is
specified using an attenuation of 0 dB.
R3765AG/67AG: ±0.5dB (50MHz, 7dBm, 23°C ± 5°C)
For OPT 10 ±0.5dB (50MHz, +2dBm, 23°C ± 5°C)
If OPT 10 is installed, the accuracy is
specified using an attenuation of 0 dB.
Flatness 2.0dBp-p (23°C ± 5°C)
For OPT 11, 13 or 14, the accuracy is specified using TEST
PORT 1.
(Continued)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3765G Series and is the answer not in the manual?

ADVANTEST R3765G Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3765G Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals