EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Page 1102

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1102User Manual 1175.6632.02 ─ 30
Parameters:
<Skew> float
Range: -500E-9 to 500E-9
Increment: 1E-12
*RST: 0
Manual operation: See "Skew" on page 445
[:SOURce<hw>]:IQ:IMPairment[:STATe] <State>
[:SOURce]:BB:IMPairment:BBMM<ch>:STATe <State>
[:SOURce]:BB:IMPairment:FADer<ch>:STATe <State>
[:SOURce]:BB:IMPairment:IQOutput<ch>:STATe <State>
[:SOURce]:BB:IMPairment:RF<ch>:STATe <State>
Activates the impairment or correction values LEAKage, QUADrature and IQRatio
for the corresponding stream.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SOURce1:BB:IMPairment:RF1:STATe ON
Activates digital impairment.
Manual operation: See "State" on page 444
[:SOURce]:BB:IMPairment:BBMM<ch>:POFFset <PhaseOffset>
[:SOURce]:BB:IMPairment:FADer<ch>:POFFset <PhaseOffset>
[:SOURce]:BB:IMPairment:IQOutput<ch>:POFFset <PhaseOffset>
[:SOURce]:BB:IMPairment:RF<ch>:POFFset <PhaseOffset>
Adds an additional phase offset after the stream mapper.
You can shift the phase at the different stages in the signal generation flow, see:
[:SOURce<hw>]:BB:POFFset on page 986
:SCONfiguration:OUTPut:MAPPing:STReam<st>:POFFset on page 921
Parameters:
<PhaseOffset> float
Range: -999.99 to 999.99
Increment: 0.01
*RST: 0
Example:
SOURce:BB:IMPairment:RF1:POFFset 0
Manual operation: See "Phase Offset" on page 446
SOURce Subsystem

Table of Contents