Remote Control Commands
R&S
®
SMW200A
1190User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:POWer:ALC:DAMPlifier <Amplifier>
Selects the driver amplifier switching state.
Parameters:
<Amplifier> OFF | ON | AUTO | FIX | ONMG
AUTO
Switches the attenuator automatically.
ON|OFF
Switches on or off the driver amplifier.
FIXed
Fixes the last setting.
ONMG
Supplies maximum level at the output.
*RST: AUTO
Example:
POW:ALC:DAMP AUTO
Provides automatic switching of the attenuator.
Manual operation: See "Driver Amplifier" on page 569
[:SOURce<hw>]:POWer:ALC:DSENsitivity <Sensitivity>
Sets the sensitivity of the ALC detector.
Parameters:
<Sensitivity> AUTO | FIXed | LOW | MEDium | HIGH
AUTO
Selects the optimum sensitivity automatically.
LOW|MEDium|HIGH
Sets either low, medium or high sensitivity.
FIXed
Fixes the internal level detector.
*RST: AUTO
Example:
POW:ALC:DSEN FIX
Manual operation: See "Detector Sensitivity" on page 569
[:SOURce<hw>]:POWer:ALC:SONCe
Activates level control for correction purposes temporarily.
Example:
POW:ALC OFF
Deactivates automatic level control at the RF output.
POW:ALC:SONC
Executes level control (once).
Usage: Event
Manual operation: See "Readjust" on page 470