Remote Control Commands
R&S
®
SMW200A
1191User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:POWer:ALC:SLEVel <SampLevel>
Sets the sample level of automatic level control (ALC).
How To: See Chapter 8.9.2.3, "How to Enable the ALC", on page 570.
Parameters:
<SampLevel> FULL | MINimum | ATTenuated
*RST: FULL
Example:
POW:ALC:SLEV MIN
Attenuates the sample level by 10 dB.
Manual operation: See "Sample Level" on page 569
[:SOURce<hw>]:POWer:ATTenuation <Attenuation>
Sets the attenuation value of the RF signal in manual mode, set with command :
OUTPut<hw>:AMODe.
Parameters:
<Attenuation> integer
Range: depends on the installed options
Increment: 4|5|6
*RST: 0
Example:
SOURce1:POWer:ATTenuation 20dB
Manual operation: See "Attenuation" on page 565
[:SOURce<hw>]:POWer:ATTenuation:DIGital <AttDigital>
Sets a relative attenuation value for the baseband signal.
Parameters:
<AttDigital> float
Range: -3.522 to 80
Increment: 1E-3
*RST: 0
Manual operation: See "Digital Attenuation" on page 467
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE <Mode>
Selects the state the attenuator is to assume if the RF signal is switched off.
Parameters:
<Mode> UNCHanged | FATTenuation
SOURce Subsystem