Remote Control Commands
R&S
®
SMW200A
1231User Manual 1175.6632.02 ─ 30
Parameters:
<Gateway> string
Range: 0.0.0.0 to ff.ff.ff.ff
Example: See Example "Retrieving information on network-related set-
tings" on page 1217.
Manual operation: See "Default Gateway" on page 788
:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK <Mask>
Sets the subnet mask.
Parameters:
<Mask> string
Example: See Example "Retrieving information on network-related set-
tings" on page 1217.
Manual operation: See "Subnet Mask" on page 787
:SYSTem:COMMunicate:SERial:BAUD <Baud>
Defines the baudrate for the serial remote control interface.
Parameters:
<Baud> 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200
*RST: n.a. (factory preset: 115200)
Example: See Example "Finding out the used VISA ressource strings"
on page 1217.
Manual operation: See "Baud Rate" on page 790
:SYSTem:COMMunicate:SERial:PARity <Parity>
Enters the parity for the serial remote control interface.
Parameters:
<Parity> NONE | ODD | EVEN
*RST: n.a. (factory preset: NONE)
Example: See Example "Finding out the used VISA ressource strings"
on page 1217.
Manual operation: See "Parity" on page 790
:SYSTem:COMMunicate:SERial:RESource?
Queries the visa resource string for the serial remote control interface. This string is
used for remote control of the instrument.
SYSTem Subsystem