Maintenance, Storage and Disposal
R&S
®
SMW200A
1280User Manual 1175.6632.02 ─ 30
17.3.3.1 Internal Adjustment Settings
Access:
► Select "System Config > Setup > General > Internal Adjustments".
In this dialog, you can perform internal calibration routines, and get information on
the last performed calibration.
The "Temperature Offset" indicates the deviation of the current temperature of the
instrument, compared to the temperature of the last adjustment.
The remote commands required to define these settings are described in Chapter 14.8,
"CALibration Subsystem", on page 863.
How to: See Chapter 17.3.3.2, "How to Use the Internal Adjustments", on page 1282.
Adjust All
Performs all available internal calibration routines of the instrument, considering all
installed paths.
Note: "Adjust All" is disabled when an R&S SZU IQ Upconverter is connected. In this
configuration, the R&S SMW provides the adjustment of the R&S SZU, see Adjust
External Frequency Converter.
How to: See Chapter 17.3.3.2, "How to Use the Internal Adjustments", on page 1282.
Remote command:
:CALibration:ALL[:MEASure]? on page 864
Last Full Adjustment
Displays the date of the last fully performed adjustment.
Remote command:
:CALibration<hw>:ALL:DATE? on page 864
Performing Maintenance Tasks