Configuring the RF Signal
R&S
®
SMW200A
497User Manual 1175.6632.02 ─ 30
Trigger sources:
The rotary knob at the front panel.
The [Arrow] keys at the front panel.
The corresponding remote control commands.
An externally applied trigger signal.
To step through the sweep frequencies or levels:
In manual mode:
Set the Current Frequency or Current Level values
Use the [Up] and [Down] keys or the rotary knob
In remote control mode:
Use the commands [:SOURce<hw>]:FREQuency:MANual or
[:SOURce<hw>]:POWer:MANual with the UP or DOWN parameter
Steps that would exceed the sweep range are ignored.
Table 8-4: Cross-reference between manual and remote control in Step / Extern Step modes (Sweep/
List)
Manual control mode:
"Step / Extern Step"
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep :TRIGger<hw>:FSWeep:SOURce SINGle [BUS] "Step" or
:TRIGger<hw>:FSWeep:SOURce EXTernal [EXTernal] for "Extern
Step"
:SOURce<hw>:SWEep:FREQuency:MODE STEP
RF level sweep :TRIGger<hw>:PSWeep:SOURce SINGle [BUS] "Step" or
:TRIGger<hw>:PSWeep:SOURce EXTernal [EXTenal] "Extern Step"
:SOURce<hw>:SWEep:POWer:MODE STEP
LF frequency sweep :TRIGger<hw>:LFFSweep:SOURce SINGle [BUS] "Step" or
:TRIGger<hw>:LFFSweep:SOURce EXTernal [EXTenal] "Extern Step"
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE STEP
List :SOURce<hw>:LIST:TRIGger:SOURce SINGle "Step" or
:SOURce<hw>:LIST:TRIGger:SOURce EXTernal "Extern Step"
:SOURce<hw>:LIST:MODE STEP
Varying the RF Signal in List or Sweep Mode