Logic analyzer (option R&S RTM-B1, MSO)
R&S
®
RTM3000
395User Manual 1335.9090.02 ─ 09
Couple Levels
Sets all threshold and hysteresis values to the values of the first nibble (D0...D3).
Remote command:
DIGital<m>:THCoupling on page 770
Set all Deskew to Zero
Aligns the skew between the probe boxes of the digital channels and the probe con-
nectors of the analog channels.
Remote command:
DIGital<m>:DESKew on page 771
Technology
Selects the threshold level: 3 predefined threshold levels and user-defined threshold
are available.
"TTL: 1.4V"
Sets the threshold to 1.4 V, which is typically used in transistor–tran-
sistor logic (TTL).
"CMOS: 2.5V"
Sets the threshold to 2.5 V, which is typically used in complementary
metal-oxide–semiconductor technology (CMOS).
"ECL: -1.3V"
Sets the threshold to -1.3 V, which is typically used in emitter-coupled
logic (ECL).
"Threshold"
Selects the user-defined threshold.
Enter the value in Threshold.
Remote command:
DIGital<m>:TECHnology on page 769
Threshold
Sets the threshold level value between -2 V and +8 V in steps of 10 mV, or shows the
value of the selected technology.
Remote command:
DIGital<m>:THReshold on page 770
Hysteresis
Defines the size of the hysteresis to avoid the change of signal states due to noise.
Remote command:
DIGital<m>:HYSTeresis on page 770
Label
Opens a menu to specify user-defined text labels for the individual logic channels.
Bit ← Label
Selects the logic channel or "Bit" for labeling.
For the pod "D7...D0", you can select bit "D0", "D1", "D2", ... or "D7".
For the pod "D15...D8", you can select bit "D8", "D9", "D10", ... or "D15".
Label ← Label
Enables or disables the user-defined label for the selected logic channel.
Logic analyzer settings