EasyManua.ls Logo

Rohde & Schwarz RTM3K-COM2

Rohde & Schwarz RTM3K-COM2
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTM3000
562User Manual 1335.9090.02 ─ 09
BPLot:REPeat............................................................................................................... 564
BPLot:RESet................................................................................................................. 564
BPLot:STATe................................................................................................................. 564
BPLot:ENABle
Enables the Bode plot diagram.
Parameters:
<PlotEnable> ON | OFF
*RST: OFF
BPLot:AUToscale
Sets the start and stop frequencies for the Bode plot automatically.
Usage:
Event
BPLot:AMPLitude:PROFile:COUNt
Sets the number of different points that you can define for the amplitude profile.
Parameters:
<NumberOfPoints> Range: 2 to 16
Increment: 1
*RST: 4
BPLot:AMPLitude:PROFile:POINt<n>:AMPLitude
Sets the amplitude of the specified point of the amplitude profile.
Parameters:
<AmplitudeMode>
BPLot:AMPLitude:PROFile:POINt<n>:FREQuency
Sets the frequency of the specified point of the amplitude profile.
Parameters:
<Frequency>
BPLot:AMPLitude:MODE
Sets the amplitude mode.
Parameters:
<AmplitudeMode> CONStant | PROFil
CONStant
In the amplitude constant mode, there is a fixed amplitude for all
frequencies.
Applications

Table of Contents

Related product manuals