EasyManua.ls Logo

Rohde & Schwarz RTM3K-COM2

Rohde & Schwarz RTM3K-COM2
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTM3000
647User Manual 1335.9090.02 ─ 09
Return values:
<AcknowledgeValue> Decimal value
Usage: Query only
BUS<b>:CAN:FRAMe<n>:CSSTate?
Returns the state of the checksum.
Suffix:
<b>
.
1..4
<n> *
Selects the frame (1...n).
Return values:
<ChecksumState> OK | UNDF
UNDF: Undefined
Usage: Query only
BUS<b>:CAN:FRAMe<n>:CSValue?
Returns the checksum value.
Suffix:
<b>
.
1..4
<n> *
Selects the frame (1...n).
Return values:
<ChecksumValue> Decimal value
Usage: Query only
BUS<b>:CAN:FRAMe<n>:DLCState?
Returns the state of the data length code.
Suffix:
<b>
.
1..4
<n> *
Selects the frame (1...n).
Return values:
<DLCState> OK | UNDF
UNDF: Undefined
Usage: Query only
Serial bus analysis