Remote control commands
R&S
®
RTM3000
673User Manual 1335.9090.02 ─ 09
BUS<b>:I2S:FCOunt?
Returns the number of decoded frames.
Suffix:
<b>
.
1..4
Return values:
<FrameCount>
Usage: Query only
TRIGger:A:I2S:CHANnel:LEFT:CONDition <Comparison>
Define the operators for comparison of the decoded data words with the specified data
words on the specified channel.
Parameters:
<Comparison> OFF | EQUal | NEQual | GTHan | LTHan | INRange | OORange
INRange: in range
OORange: out of range
OFF
No range is defined.
EQUal | NEQual | GTHan | LTHan
Equal, not equal, greater than, less than. These conditions
require one data word to be set with TRIGger:A:I2S:
CHANnel:LEFT:DMIN.
TRIGger:A:I2S:CHANnel:LEFT:DMAX <MaximumValue>
Specifies the maximum data value to be found, if TRIGger:A:I2S:CHANnel:LEFT:
CONDition is set to INRange or OORange
Parameters:
<MaximumValue> The data format is decimal. The maximum value is limited by the
word length. Consider that audio words are signed numbers in
2's complement format. For example, an 8-bit data word has a
value range from -128 to 127.
TRIGger:A:I2S:CHANnel:LEFT:DMIN <MinimumValue>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<MinimumValue>