EasyManua.ls Logo

Rohde & Schwarz RTM3K-COM2

Rohde & Schwarz RTM3K-COM2
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTM3000
684User Manual 1335.9090.02 ─ 09
Parameters:
<ErrorEnable> ON | OFF
*RST: ON
TRIGger:A:MILStd:ERRor:PARity <ErrorEnable>
Enables/disables triggering when the parity is even.
Parameters:
<ErrorEnable> ON | OFF
*RST: ON
TRIGger:A:MILStd:ERRor:SYNC <ErrorEnable>
Enables/disables triggering when a sync impulse doesn't fulfil the technical require-
ments or when the transmission is not valid.
Parameters:
<ErrorEnable> ON | OFF
*RST: ON
TRIGger:A:MILStd:ERRor:TIMeout <ErrorEnable>
Enables/ disables triggering when the timeout is out of the set range.
Parameters:
<ErrorEnable> ON | OFF
*RST: ON
TRIGger:A:MILStd:FRAMe <Frame>
Parameters:
<Frame> COMMand | STATus | DATA | ALL
*RST: CST
TRIGger:A:MILStd:MCODe:CODE <ModeCode>
Sets a function for the mode code to be triggered on.
Parameters:
<ModeCode> DBControl | TSYNchronize | TSTatus | ISELftest | TSHutdown |
OTSHutdown | ITERminal | OITerminal | RESet | VECTor |
RSYNchronize | TLAStcmmand | BITWord | STSHutdown |
OSTShutdown | ANY
DBControl: dynamic bus control.
TSYNchronize: synchronize without data.
TSTatus: transmit status word.
ISELftest: initiate self-test.
Serial bus analysis

Table of Contents