Remote control commands
R&S
®
RTP
1192User Manual 1337.9952.02 ─ 12
Parameters:
<SampleRate> Range: 20 Gsample/s, R&S RTP with serial number
103000 in interleaved mode (2 channels):
40 Gsample/s
Increment: 1
*RST: 20 Gsample/s
Default unit: Sa/s
Usage: Asynchronous command
ACQuire:SRReal <RealSampleRate>
Sets the number of captured waveform points per second. It considers the samples of
the ADC, and the reduction of waveform points by decimation.
If interpolation is not active, the sample rate is the reciprocal value of the resolution
and thus also depends on the acquisition time and the record length.
If interpolation is active, the sample rate is limited to the ADC sample rate.
Parameters:
<RealSampleRate> Range: 20 Gsample/s, R&S RTP with serial number
103000 in interleaved mode (2 channels):
40 Gsample/s
Increment: 1
*RST: 20 Gsample/s
Default unit: Sa/s
Usage: Asynchronous command
Firmware/software: FW 3.40
ACQuire:RESolution <Resolution>
Indicates the time between two waveform points in the record.
Parameters:
<Resolution> A fine resolution with low values produces a more precise wave-
form record.
Range: 1E-15 to 0.5
Increment: 10E-12
*RST: 100E-12
Default unit: s
Usage: Asynchronous command
ACQuire:POINts[:VALue] <RecordLength>
Indicates the record length, the number of recorded waveform points that build the
waveform across the acquisition time. [:VALue] can be omitted.
Acquisition and setup