EasyManua.ls Logo

Rohde & Schwarz RTP044 - Reference Waveforms

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1327User Manual 1337.9952.02 ─ 12
Setting parameters:
<DiagramName> Name of the diagram on which the zoom area is based.
<ZoomName> Name of the zoom diagram
Example: See Chapter 22.3.1.2, "Creating zoom diagrams", on page 1127
Usage: Setting only
Asynchronous command
22.10.2 Reference waveforms
Reference............................................................................................................1327
Scaling................................................................................................................ 1329
Waveform data export.........................................................................................1334
Import of multichannel waveform data................................................................ 1335
22.10.2.1 Reference
REFCurve<m>:SOURce...............................................................................................1327
REFCurve<m>:STATe...................................................................................................1328
REFCurve<m>:NAME.................................................................................................. 1328
REFCurve<m>:OPEN...................................................................................................1328
REFCurve<m>:UPDate................................................................................................ 1328
REFCurve<m>:SAVE................................................................................................... 1329
REFCurve<m>:DELete.................................................................................................1329
REFCurve<m>:CLEar...................................................................................................1329
REFCurve<m>:SOURce <Source>
Selects the source waveform to be used as a reference.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<Source> C1W1 | C1W2 | C1W3 | C2W1 | C2W2 | C2W3 | C3W1 | C3W2 |
C3W3 | C4W1 | C4W2 | C4W3 | M1 | M2 | M3 | M4 | R1 | R2 |
R3 | R4 | TRK1 | TRK2 | TRK3 | TRK4 | TRK5 | TRK6 | TRK7 |
TRK8 | Z1V1 | Z1V2 | Z1V3 | Z1V4 | Z1I1 | Z1I2 | Z1I3 | Z1I4 |
Z2V1 | Z2V2 | Z2V3 | Z2V4 | Z2I1 | Z2I2 | Z2I3 | Z2I4 | DIFF1 |
DIFF2 | COMMON1 | COMMON2
Source of the reference waveform, see Chapter 22.4.2, "Wave-
form parameter", on page 1145
*RST: C1W1
Usage: Asynchronous command