Remote control commands
R&S
®
RTP
1269User Manual 1337.9952.02 ─ 12
22.9.9 Slew rate trigger
TRIGger<m>:SLEW:SLOPe..........................................................................................1269
TRIGger<m>:LEVel<n>:SLEW:LOWer........................................................................... 1269
TRIGger<m>:LEVel<n>:SLEW:UPPer............................................................................1269
TRIGger<m>:SLEW:RANGe......................................................................................... 1270
TRIGger<m>:SLEW:RATE............................................................................................1270
TRIGger<m>:SLEW:DELTa...........................................................................................1271
TRIGger<m>:SLEW:SLOPe <Slope>
Selects the edge type for the trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Slope> POSitive | NEGative | EITHer
See Chapter 22.4.3, "Slope parameter", on page 1147.
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:SLEW:LOWer <Level>
TRIGger<m>:LEVel<n>:SLEW:UPPer <Level>
Set the lower and upper voltage thresholds. When the signal crosses a level, the slew
rate measurement starts or stops depending on the selected slope.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
<n> 1..31
Indicates the trigger source:
1...4 = channel 1...4
5...27 = not available
28...31 = differential signals Diff1, Diff2 and Common1, Com-
mon2
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: Lower = -0.1, upper =0.1
Default unit: V
Usage: Asynchronous command
Trigger