Remote control commands
R&S
®
RTP
1526User Manual 1337.9952.02 ─ 12
MMEMory:AUTonaming:TEXT <NameString>
Defines a text, that can be included in the autonaming pattern.
Parameters:
<NameString> String parameter
MMEMory:AUTonaming:DEFaultpath <Path>
Sets the path where data and settings files will be stored. The factory default path is:
●
"C:\Users\Public\Documents\Rohde-Schwarz\RTx" if no USB flash drive
is connected
●
Drive letter of the USB flash drive, for example, "E:\" or "F:\" if a USB flash
drive is connected.
Parameters:
<Path> String parameter
MMEMory:AUTonaming:RESPath
Resets the path for file operations to the factory default path.
Usage:
Event
MMEMory:AUTonaming:RESall
Resets all autonaming settings to the default value, including the path.
Usage:
Event
22.16.4 Waveform data transmission
The R&S RTP provides specific data export commands for the various waveform
types. The commands transmit the data of the waveform points from the instrument to
the controlling computer. The data can be used in MATLAB, for example.
The commands are described in the relevant chapters:
●
Analog waveforms: Chapter 22.8.6, "Waveform data", on page 1204
●
Reference waveforms: Chapter 22.10.2.3, "Waveform data export", on page 1334
●
Math waveforms: Chapter 22.10.3, "Mathematics", on page 1336
●
Spectrum waveforms: Chapter 22.13.2, "Waveform data", on page 1455
●
Logic channels: Chapter 22.18.5, "MSO data", on page 2217
●
I/Q data: Chapter 22.21.2, "I/Q data output", on page 2254
Data management